Basics Of Calibration

Calibration – The Definition

Calibration is the comparsion of measurement device or an instrument(device under test, DUT) against a known with equal or better standard. The standard in a measurement is considered to be more correct of the two and one would calibrate the device under test to know far it deviates from the standard.
Usual calibration usually done by commercial calibration laboratories uses a manufacturer’s calibration procedure and is performed with a reference standard multiple times more accurate typically at four times with accuracy of the DUT.

So, Why Do We Need to Calibrate?

Having instruments that are calibrated insures quality output products. Out of tolerance (OOT) instruments tend to give wrong readings resulting in unnecessary back jobs and process failures. Bad or low quality products would pass as good ones resulting in warranty costs, and good products as non-conformance to quality ones, resulting to unnecessary reworks. So basically, calibration is all about producing high quality which is equals to money.

Common calibration terms

Out of Tolerance Conditions:

If the results are outside of the instrument’s performance specifications it is considered an OOT (Out of Tolerance) condition and will result in the need to adjust the instrument back into specification.

Optimization:

Adjusting a measuring instrument to make it more accurate is NOT part of a “Typical” calibration and is frequently referred to as “Optimizing” or “Nominalizing” an instrument. (this is a common misconception) Only reputable and experienced calibration providers should be trusted to make adjustments on critical test equipment.

As Found Data:

The reading of the instrument before it is adjusted.
delays.

As Left Data:

The reading of the instrument after adjustment or “Same As Found” if no adjustment was made.

Without Data:

Most calibration labs charge more to provide the certificate with data and will offer a “No-Data” option. In any case “As-Found” data must be provided for any OOT condition.

Limited Calibration:

Sometimes certain functions of an instrument may not be needed by the user. It may be more cost effective to have a limited calibration performed (This can even include a reduced accuracy calibration).

TUR – Test Uncertainty Ratio:

The ratio of the accuracy of the instrument under test compared to the accuracy of the reference standard.

ISO 9000 Calibration

ISO 9000 calibrations are crucial for many industries. The following is required for ISO 9000 Compliant Calibrations.

An Accredited Calibration Lab Performing the Work:

The calibration laboratory employed to perform the calibration must be an ISO 9001:2000 accredited lab or be the original equipment manufacture.

Documented Calibration Procedures:

It is critical that a valid calibration procedure be used based on the manufacture’s recommendations and covering all aspects of the instrument under test.
Trained Technicians: Proper Training must be documented for each discipline involved in performing the calibration.
Traceable Assets: The calibration provider must be able to demonstrate an unbroken chain of traceability back to NIST.

Proper Documentation:

All critical aspects of the calibration must be properly documented for the certificate to be recognized by an ISO auditor.

A Comprehensive Equipment List:

For any manufacture to pass an ISO audit regarding calibration they must demonstrate that they have a comprehensive equipment list with controls in place for additions, subtractions and custodianship of equipment.

Calibrated and NCR Items Properly Identified:

The equipment list must identify any units that do not require calibration and controls must be in place to ensure that these units are not used in an application that will require calibration.

A Proper Recall System:

A procedure should be established with time frames for recall notification, an escalation procedure, and provisions for due-date extension.

Equipment Custodianship:

Responsibilities for ensuring the equipment is returned to the cal lab should be assigned and delegated.

An OOT Investigation Log:

Any instrument found out of tolerance requires that an investigation be performed to determine the impact on manufacturing. Records and reports need to be maintained.

ISO/IEC 17025 Calibration

ISO/IEC 17025 Calibration:

As a general rule 17025 calibrations are required by anyone supplying the automotive industry and it has also been voluntarily adopted by numerous companies in FDA regulated industries.
ISO/IEC 17025 is an international standard that assesses the technical competency of calibration laboratories. ISO/IEC 17025 covers every aspect of laboratory management, ranging from testing proficiency to record keeping and reports. It goes several steps beyond a ISO 9001:2000 certification.
A “17025” calibration is a premium option that provides additional information about the quality of each measurement made during the calibration process by individually stating the uncertainty calculation of each test point.

When Do We Need To Calibrate?

How Calibration Intervals are Determined
Calibration intervals are to be determined by the instrument “owner” based on manufacture recommendations. Commercial calibration laboratories can suggest intervals but in most cases they are not familiar with the details of the instrument’s application.
The OEM intervals are typically based on parameters like mean drift rates for the various components within the instrument. However, when determining calibration intervals as an instrument “owner” several other factors should be taken into consideration such as: the required accuracy vs. the instrument’s accuracy, the impact an OOT will have on the process, and the performance history of the particular instrument in your application.

How to Implement or Improve a Calibration Program ?

Any successful calibration program must begin with an accurate recall list of your test, measurement and diagnostic equipment.
The recall list should contain a unique identifier which can be used to track the instrument, the location, and the instrument’s custodian (Often asset management software, bar-coding systems, and physical inventories are used to help establish accurate recall lists).
It is important when assembling a recall list that modules, plug-ins, and small handheld tools are not overlooked. Also, you may have several “home-made” measuring devices (e.g. Test Fixtures) which will also need to be captured on your equipment list for a reliable calibration program.
The next step is to identify all of the instruments on your recall list which may not require calibration due to redundancies in your testing process (A commercial calibration laboratory should be able to aid you in identifying these instruments).
After creating an accurate recall list procedures must be established for adding new instruments, removing old or disposed instruments, or making changes in instrument custodianship. Recall reports should be run with sufficient time for both the end user and the service provider to have the unit calibrated with a minimal impact on production.
A late report identifying any units about to expire or already expired will ensure 100% conformity. A full service calibration laboratory will supply these recall reports and will provide special escalation reporting when equipment is not returned for service.
(Some calibration labs offer the choice of web-based equipment management systems that allow their customer to perform recall reports, late reports and keep electronic versions of their calibration certificates.)

Avoiding Production Delays

Obtain timely equipment calibrations without shutting down a line for days.
Look for a calibration service provider that can perform onsite (or in-place) calibrations at your facility. Often when your volume is more than 20 calibrations, scheduling onsite calibration saves time and lowers cost.

Make sure you find a “one-source”

calibration provider that has sufficient capabilities to calibrate nearly all of your equipment during the onsite, reducing the delays and the expense of using an additional subcontractor.
Other options for reducing downtime include mobile Calibration lab services, scheduled depot calibrations, calibrations during shutdowns, scheduled pick-up and delivery, and weekend or nightshift calibrations.

Should We Calibrate Ourselves?

Most companies discover they cannot effectively perform their own calibrations for many reasons. The most frequent issues with internal calibrations are:

Cost of standards:

Often, the cost of the assets with the required accuracy to perform the calibration is prohibitive (It could take years of calibrations to pay for one standard).

Developing Procedures:

Many manufacture’s procedures are not readily available. Sometimes they require research and development. This can cost hundreds of hours of labor.

Productivity of Technicians:

Often a non-commercial calibration laboratory’s productivity per employee is only a fraction of what can be obtained through an external commercial calibration laboratory who specializes in automation, efficient procedures and experienced management.

Cost of Management:

Managing the employees, assets, maintenance and processes of a calibration lab can be burdensome on existing management staff.
Not a core competency: The overall management burden of the operation distracts from the core competency of the company.

Comments

Popular posts from this blog

BASICS OF TRIPS, INTERLOCKS, PERMISSIVES & SEQUENCES

■ TRIP : The term trip refers to an action that is initiated by the control system and which forces a device or devices to a pre-determined state. Example of Trip Signals: Close Valve, Open Valve, Stop motor, etc. The Safety Instrument System (SIS) or a Hardwired systems normally initiate trips, however the PLCs or DCS may also initiate trips provided the necessary independence and SIL ratings are met.Once a device or devices have been forced to a pre-determined state by the action of a Trip they will remain in that state until the Trip is manually reset by a conscious operator action. For example: High level in a vessel initiates a trip system which stops the pump feeding that vessel, the pump will remain stopped even if the level in the vessel falls to a safe level. The Trip must be ‘reset’ by the operator before the pump can be re-started. The Trip can only be ‘reset’ if the level in the tank has fallen to a safe level. Resetting the Trip will no...

Instrument Range Selection

Instrument range shall be selected based on operating range of process to be measured. In common practice, it is recommended to select instrument with full scale range that can give operating process reading between 30% and 70% especially for pressure gauge. The following is an example of selecting pressure gauge range Given Process data: Minimum operating pressure: 5 barg Maximum operating pressure: 9.5 barg Maximum Design pressure: 25 barg Available standard Pressure Gauge range: 0 – 10 barg, 0 – 16 barg, 0 – 25 barg [X] Selecting 0-10 barg will give 95% reading during maximum operating condition. Instrument has only narrow margin above maximum operating pressure. [√] Selecting 0-16 barg range will give best reading during operating condition. However, the maximum overpressure of pressure gauge is commonly 1.3x of the maximum scale which is only 16×1.3 = 20.8 barg. In such case, pressure gauge shall be equipped with overrange protector to withstand ...

What is difference gap control ?

Differential gap control is similar to on off control except that a band or gap exists around the control point. Use:  In industry differential gap control is often found in non critical level control applications where it is desirable only to prevent a tank from flooding or drying. When a measured variable exceeds the upper gap the control valve will open fully or be closed fully. Similarly when it exceeds the lower gap it will open or close fully.